- Chemical
Storage back
to top
S |
Incompatible Chemicals Stored Together
- Acids/bases
- Flammables/oxidizers
- Organic/inorganic Acids
- Water Reactives/Water or Water-based Compounds
- Oxidizers Stored on Incompatible Shelf Material
|
Example Example |
| Containers Not Sealed Properly |
Example |
| Containers Compromised
|
Example |
O |
Fume hood utilized for storage while actively being utilized
for chemical operations |
|
F |
|
|
- CHP Manual
back to top
| S |
None |
Request |
| O |
Not Completed |
Example |
Varying degrees of incomplete
- No SOPs for Select Carcinogens, Reproductive Toxins and
Acutely Toxic Chemicals
- No Chemical Inventory
- No Lab Specific Training Documentation
- Incomplete ID Page
- Information Not Current
|
Example |
| F |
|
|
- Compressed
Gas Cylinders back to top
S |
Unsecured |
Example |
| Not secured properly |
|
| Exceeding limits for storage per UK policy |
Fact Sheet |
Toxic gases not in continuously ventilated hood
or gas cabinet1. Would include but not be limited
to:
arsine, diborane, germane, phosphine, nitric oxide, methyl bromide,
boron trifluoride, chlorine, chlorine trifluoride, dichlorosilane,
hydrogen fluoride, nitrogen dioxide, phosgene, sulfur tetrafluoride,
ammonia, boron trichloride, boron trifluoride, carbon monoxide,
carbonyl sulfide, ethyl chloride, hydrogen bromide, hydrogen
chloride, hydrogen sulfide, silane, and disilane |
Example |
Incompatible gases stored together • Flammables/oxidizers |
|
| Utilizing regulator as isolation device |
Example |
O |
Valve caps not on cylinders in storage |
Example |
F |
|
|
- Controlled
Access back to top
| S |
|
|
| O |
Lab left unlocked and unattended |
|
| Children in lab |
Example |
| Pets in lab |
|
| F |
|
|
- Door
Signage back to top
| S |
|
|
| O |
None |
Forms |
| Incomplete |
|
| Outdated/incorrect information |
|
| F |
|
|
- Electrical
back to top
| S |
Damaged/frayed power cords |
|
Use of appliances not UL listed for application •
Blenders • Heat guns/hair dryers |
|
| O |
Ext cords utilized for permanent wiring |
Example |
| Use of multiple power strips inline |
Example |
| No strain relief on energized cords |
Example |
F |
|
|
- Eyewash
back to top
| S |
Blocked/obstructed eyewash |
Example |
| O |
|
|
| F |
Non-compliant eyewash |
Example |
| No eyewash |
|
- Fire
Extinguisher back to top
| S |
No Fire extinguisher |
E-mail |
| Fire ext. discharged and not reported |
|
| Fire ext. blocked |
Example |
| O |
Fire ext. not in wall mount |
|
| F |
Fire ext. not inspected annually |
|
| Fire ext. not charged – “not in
the green” |
Example |
| Fire ext. not mounted |
|
- Flammable
Storage back to top
S |
Storage amounts exceed Solvent Storage Policy |
Policy |
| Flammables stored in unapproved refrigerator |
|
Unapproved flammable storage cabinet • Three latch
inoperable • Not FM or UL listed |
|
| Cabinet not closed |
Example |
| Vent caps removed |
|
O |
|
|
F |
|
|
- Food
back to top
S |
Evidence consistent with eating and/or drinking
in the lab |
Example |
| Storage of food in lab area |
Example |
O |
|
|
F |
|
|
- Fume
Hood back to top
| S |
Sash above working height during use |
Example |
| Alarm rendered inoperable via tampering |
|
| Using hood when failing |
Example |
| Baffles obstructed |
Example |
Incompatible chemical utilized with standard
fume hood
• Perchloric acid |
Policy |
| O |
Excessive chemicals/equipment in hood |
Example |
| F |
Hood in alarm mode |
E-mail |
| Alarm not functioning |
E-mail |
| No flow indicator and/or alarm |
Guide |
- Hazardous
Waste back to top
S |
No label |
|
Label incomplete • “Hazardous Waste”
not on label • No date as to when full •
No name of contents listed |
Example |
| Waste not ticketed for pick-up when container full |
E-mail |
| Open containers of HW |
Example |
| Evidence of improper disposal |
|
O |
|
|
F |
|
|
- Housekeeping
back to top
S |
Means of egress, i.e., aisles, doorways blocked |
|
O |
Chemical stored in aisle ways – obstructing
egress and spill potential |
|
| Slip/Trip hazards – power and extension
cords, liquids on floor |
Example |
| Overabundance of combustibles |
|
F |
|
|
- Labeling
back to top
S |
Chemical containers not labeled |
Example |
| Illegible container labels |
Example |
Label incomplete
• No chemical name |
|
O |
Food stuffs utilized for research not labeled
for intended use, i.e., “food not to be used for human
consumption” |
|
F |
|
|
- Other
back to top
| S |
Machine not guarded |
|
| O |
No vacuum trap utilized with vacuum source |
|
| Utilizing chipped or broken glassware |
|
| Improper disposal of glassware (deposited in regular trash
in lab) |
|
| Overfilled sharps container |
Example |
| No annual certification of biological safety cabinet |
Example |
F |
|
|
- Peroxide
Former back to top
S |
Not dated for disposal in accordance with guide sheet |
Policy |
| Not disposed of by mfg’er expiration date |
Example |
O |
|
|
F |
|
|
- PPE
back to top
| S |
Not wearing PPE in accordance with CHP PPE Hazard Assessment |
Request
|
Improper storage • Contamination of PPE
• Degradation of PPE |
Example |
| Improper PPE selected |
|
Improper use of PPE • Improper type •
Wearing gloves outside of lab |
|
| O |
Improper attire |
|
F |
|
|
- Safety
Shower back to top
S |
Blocked/obstructed |
Example |
| Shower activation handle tied back |
Example |
O |
|
|
F |
No shower |
|
Non-compliant shower
• No stay open valve |
|
| Handle height greater than 69 |
|
- Training
back to top
S |
|
|
O |
All affected employees not received Chemical Hygiene Plan/Laboratory
Safety Training |
Training Page |
| All affected employees not received Lab Specific Training
|
Training
Forms |
F |
|
|
|